9. High and Low value Adjustment
-Water goes up and turn the sample pressure calibrator

HMK-22 Fisher sub sieve sizer and Fisher Model 95 parts

HMK-22 Fisher sub sieve sizer and Fisher Model 95 parts

-Check the high value on the sub sieve sizer calibrator sticker,if the high value is x.x such as 6.4,make the red needle point to 6.4
-If water level is higher than 6.4,adjust high needle valve anti-clockwise until the water level becomes stable
-Turn the range control to II and the water level comes down
-Check the low value on the sub sieve sizer calibrator sticker,if it is x.x such as 2.7 point the red needle to 2.7
-If the water level is too high,adjust low needle valve anti-clockwise until the water level becomes stable
-Turn the range control back to I
-Commissioning work is finished,it is ready for testing powder