Particle Size Analyzers

HMK-22 Fisher Sub Sieve Sizer – 6 – Install Sample Pressure Calibrator And Test It

HMK-22 Fisher Sub Sieve Sizer – 6 – Install Sample Pressure Calibrator And Test It

6.Install Sample Pressure Calibrator and Test It
-Dismantle the knob on rack and pinion control
-Insert the hex connector
-Screw it tightly
-Insert the sample pressure calibrator
-Put the pressure measurement tube on the base block assembly
-There is a red line,press it tight until reach 50
-Check the dial
-If it stays at a number such as 14,then later we should press the powder to 14
-Parallel the needle to the baseline and point it to 0.75 and check if the water is enough

Leave a Reply

X